Commissioning
Deltabar FMD72
60
Write permission
Operators/Service engineers/Expert
Description
Selection of data packages for up/download function in Fieldcare and PDM.
Prerequisite
DIP switch set to "SW" and "Damping" set to "on". If you download using the factory setting
"Configuration copy", all parameters required for a measurement will be downloaded. The
functionality of the "Electronics replace" setting is reserved for Hauser Service and can be
accessed only if the correct device access code is entered.
Options
• Configuration copy: This option overwrites general configuration parameters with the exception
of the serial number, order number, calibration, pos. zero adjust, application and day information.
• Device replacement: This option overwrites general configuration parameters with the exception
of the serial number, order number, calibration and pos. zero adjust.
• Electronics replace: This option overwrites general configuration parameters.
Factory setting
Configuration copy
11.11 Configuring the local display
11.11.1 Adjusting the onsite display
The onsite display can be adjusted in the following menus:
Setup
®
Advanced Setup
®
Display
11.12 Protecting settings from unauthorized access
The settings can be protected from unauthorized access in two ways:
• Locking via write protection switch (hardware locking)
• Locking via parameter (software locking)