Commissioning
Deltabar FMD72
46
The values entered for "Empty Calib./Full Calib.", "Set LRV/Set URV", and the applied
pressures must be at least 1% apart. The value will be rejected, and a warning message displayed,
if the values are too close together. Further limit values are not checked, i.e. the values entered
must be appropriate for the sensor module and the measuring task for the device to be able to
measure correctly.
Description
1
Carrying out "Position adjustment"(
0 mbar
300 mbar
3 m
0 m
1
2
A0017655
1
See table, step 7.
2
See table, step 8.
2
Select the "Level" measuring mode via the
"Measuring mode (005)" parameter.
Menu path: Setup
®
Measuring mode
3
Select a pressure engineering unit via the "Press Eng.
Unit (125)" parameter, here "mbar" for example.
Menu path: Setup
®
Press Eng. Unit
4
Select the "In pressure" level mode via the "Level
selection (024)" parameter.
Menu path: Setup
®
Extended Setup
®
Level
®
Level Selection.
5
Select a level unit via the "Unit before lin. (025)"
parameter, here "m" for example.
Menu path: Setup
®
Extended Setup
®
Level
®
Unit Before Lin.
3
0
h
[m]
0
300
p
[mbar]
20
4
I
[mA]
0
3
h
[m]
A0017658
1
See table, step 7.
2
See table, step 8.
3
See table, step 10.
4
See table, step 11.
6
Select the "Wet" option via the "Calibration mode
(027)" parameter.
Menu path: Setup
®
Extended Setup
®
Level
®
Calibration Mode
7
The container is filled to the lower calibration point.
The pressure here is, for example, 0 mbar (0 psi).
Select "Empty Calib. (028)" parameter.
Menu path: Setup
®
Extended Setup
®
Level
®
Empty Calibration
Enter the level value, here 0 m for example. The
pressure value present is assigned to the lower level
value by confirming the value.
8
The container is filled to the upper calibration point.
The pressure here is, for example, 300 mbar (4.35
psi).
Select the "Full Calib. (031)" parameter.
Menu path: Setup
®
Extended Setup
®
Level
®
Full Calibration
Enter the level value, here 3 m (9.8 ft) for example.
The pressure value present is assigned to the upper
level value by confirming the value.
9
If calibration is performed with a medium other than
the process medium, enter the density of the
calibration medium in "Adjust density (034)".
Menu path: Setup
®
Extended Setup
®
Level
®
Adjust Density
10
Use the "Set LRV (166)" parameter to set the level
value for the lower current value (4 mA) (0 m (0 ft)).
Menu path: Setup
®
Extended Setup
®
Current
Output
®
Set LRV