Description of device parameters
Deltabar FMD72
84
Navigation
Setup
®
Advanced Setup
®
Level
®
Empty Height (030)/(186)
Write permission
Operators/Service engineers/Expert
Description
Enter the height value for the lower calibration point (container empty). The unit is selected via the
"Height unit (026)" parameter.
Prerequisite
• "Level Selection" = "In height"
• "Calibration mode" = Dry -> entry
• "Calibration mode" = Wet -> display
Factory setting
0,0
Full Calib. (031)/(012)
Navigation
Setup
®
Advanced Setup
®
Level
®
Full Calib. (031)/(012)
Write permission
Operators/Service engineers/Expert
Description
Enter the output value for the upper calibration point (container full). The unit defined in "Unit
Before Lin." must be used.
Note
• In the case of wet calibration, the level (container full) must actually be available. The associated
pressure is then automatically recorded by the device.
• In the case of dry calibration, the level (container full) does not have to be available. The associated
pressure has to be entered in the "Full Pressure" parameter for the "In pressure" level selection.
The associated height has to be entered in the "Full height" parameter for the "In height" level
selection.
Factory setting
100,0
Full Pressure (032)/(187)
Navigation
Setup
®
Advanced Setup
®
Level
®
Full Pressure (032)/(187)
Write permission
Operators/Service engineers/Expert
Description
Enter the pressure value for the upper calibration point (container full). See also "Full Calib.".
Prerequisite
• "Level Selection" = In pressure
• "Calibration mode" = Dry -> entry
• "Calibration mode" = Wet -> display
Factory setting
URL of the sensor module
Full height (033)/(188)
Содержание Deltabar FMD72
Страница 2: ......
Страница 107: ......
Страница 108: ...www endress com worldwide 71191239 71191239 BA01044P 00 EN 02 12 71191239 CCS COSIMA...