Technical data
Cerabar PMP51B Analog
50
Hauser
Vibration resistance
Single compartment housing
Mechanical construction
Sine wave oscillation IEC 61298-3:2008 Shock
Device
10 Hz to 60 Hz: ±0.35 mm (0.0138 in)
60 Hz to 1000 Hz: 5 g
30 g
Device with a "compact" or "temperature isolator" type of
diaphragm seal
10 Hz to 60 Hz: ±0.15 mm (0.0059 in)
60 Hz to 1000 Hz: 2 g
30 g
1)
Use either a device with a temperature isolator or with a capillary in applications with very high
temperatures. If vibrations additionally occur in the application, Hauser recommends the use of a
device with a capillary. If a device with a temperature isolator or capillary is used, the device must be
mounted with a mounting bracket.
Electromagnetic
compatibility (EMC)
• Electromagnetic compatibility as per EN 61326 series and NAMUR recommendation
EMC (NE21)
• With regard to the safety function (SIL), the requirements of EN 61326-3-x are satisfied
• Maximum deviation with interference influence: < 0.5% of span with full measuring
range (TD 1:1)
For more details refer to the EU Declaration of Conformity.