Mounting
Cerabar PMP51B Analog
18
Hauser
5.2.3
Pressure measurement in liquids
3
1
2
A0038732
1
Device
2
Shutoff device
3
Water pocket pipe
Mount the device with the shutoff device below or at the same level as the tapping point.
5.2.4
Level measurement
A0038733
• Always mount the device below the lowest measuring point
• Do not install the device at the following positions:
• In the filling curtain
• In the tank outlet
• In the suction area of a pump
• At a point in the tank which could be affected by pressure pulses from the agitator
• Mount the device downstream from a shutoff device: the calibration and function check
can be carried out more easily
5.2.5
Thermal insulation with diaphragm seal directly mounted
The device may only be insulated up to a certain height. The maximum permitted
insulation height is indicated on the device and applies to an insulation material with a
heat conductivity ≤ 0.04 W/(m x K) and to the maximum permitted ambient and process
temperature. The data were determined under the most critical application "quiescent air".
Maximum permitted insulation height, indicated here on a device with a flange: