Product design
3.1.1
Metal process membrane
Standard device (without diaphragm seal)
1
2
3
p
A0043089
1
Measuring element
2
Channel with fill fluid
3
Metal process membrane
p
Pressure
The pressure deflects the metal process membrane of the sensor. A fill fluid transfers the
pressure to a Wheatstone bridge (semi-conductor technology). The pressure-dependent
change in the bridge output voltage is measured and evaluated.
Advantages:
• Can be used for high process pressures
• High long-term stability
• High overload resistance
• Secondary containment for enhanced integrity
• Significantly reduced thermal effect e.g. compared to diaphragm seal systems with
capillaries