Cerabar PMP51B Analog
Electrical connection
Hauser
29
Devices with M12 plug
2
1
3
4
A0011175
1
2
Not assigned
3
Signal –
4
Ground
6.3
Ensuring the degree of protection
6.3.1
Cable entries
• Gland M20, plastic, IP66/68 TYPE 4X/6P
• Gland M20, brass nickel plated, IP66/68 TYPE 4X/6P
• Gland M20, 316L, IP66/68 TYPE 4X/6P
• Thread M20, IP66/68 TYPE 4X/6P
• Thread G1/2, IP66/68 TYPE 4X/6P
If the G1/2 thread is selected, the device is delivered with an M20 thread as standard
and a G1/2 adapter is included with the delivery, along with the corresponding
documentation
• Thread NPT1/2, IP66/68 TYPE 4X/6P
• Dummy plug transport protection: IP22, TYPE 2
• *Cable 5 m, IP66/68 TYPE 4X/6P pressure compensation via cable
• *Valve plug ISO4400 M16, IP65 TYPE 4X
• Plug HAN7D, 90 deg. IP65 NEMA Type 4X
• Plug M12
When housing is closed and connecting cable is plugged in: IP66/67, NEMA Type 4X
When housing is open or connecting cable is not plugged in: IP20, NEMA Type 1
NOTICE
Plug M12 and plug HAN7D: incorrect mounting can invalidate the IP protection class!
‣
The degree of protection only applies if the connecting cable used is plugged in and
screwed tight.
‣
The degree of protection only applies if the connecting cable used is specified according
to IP67, NEMA Type 4X.
‣
The IP protection classes are only maintained if the dummy cap is used or the cable is
connected.
6.4
Post-connection check
After wiring the device, perform the following checks:
Is the potential matching line connected?
Is the terminal assignment correct?
Are the cable glands and dummy plugs leak-tight?
Are the fieldbus connectors properly secured?
Are the covers screwed down correctly?