Commissioning
Cerabar PMP51B Analog
32
Hauser
Setting the upper range value (pressure or scaled variable)
1. Desired pressure for upper range value is present at device.
2. Press the "Span" key for at least 3 s.
3. When the LED lights up briefly, the pressure present has been accepted for the upper
range value.
4. Does the LED on the electronic insert not light up?
Applied pressure has not been accepted for upper range value.
Wet calibration is not possible if the
Scaled variable
option has been selected in
the
Assign PV
parameter and the
Table
option has been selected in the
Scaled
variable transfer function
parameter.
Checking the settings (pressure or scaled variable)
1. Press the "Zero" key briefly in order to display the lower range value.
2. Press the "Span" key briefly in order to display the upper range value.
3. Press the "Zero" and "Span" keys simultaneously in order to display the calibration
offset.
Resetting the device
‣
Press and hold the "Zero" and "Span" keys simultaneously for at least 12 s.