Product description
Cerabar PMP51B Analog
12
Hauser
3.1.2
Level measurement (level, volume and mass)
Standard device (without diaphragm seal)
h =
p
ρ g
h
A0038343
h
Height (level)
p
Pressure
ρ
Density of the medium
g
Gravitational constant
Device with diaphragm seal
h =
p
ρ g
h
A0038342
1
Sample illustration: diaphragm seal with capillary
h
Height (level)
p
Pressure
ρ
Density of the medium
g
Gravitational constant
Advantages:
• Volume and mass measurements in any vessel shape with a freely programmable
characteristic curve
• Has a wide range of uses, e.g.
• For foam formation
• In vessels with agitators or screen fittings
• For liquid gases