Instruction Manual
IM-106-880C, OI
Fault Isolation
April 2017
Fault
Isolation
17
3
Cal Warning (Calibration
warning) Cal Failed (Calibration
failed) Probable Cause
Recommended Corrective
Action
Low sample gas flow in sensor housing due Check the following portions of the flow path for plugging:
to flow path plugging
blowback filter
in-situ filter
eductor outlet path
COe sensor degraded
•
COe Slope Error (Slope not
between 200 ppm/Ohm to
4,500 ppm/Ohm)
Replace COe sensor or increase warning level (SYSTEM, CALIB SETUP, COe Slope
Warn via HART.)
•
•
COe Constant Error (Constant
not between -100 Ohms to +100
Ohms)
Calibration Gas concentration incorrect or
gas bottle empty
O
2
TOL Error (not within 10% of configured O2 test gas).
COe TOL Error (not within 30% of configured CO test gas).
Table continued on next page
Figure 8-4. Oxygen and Cell Output
Содержание Rosemount OCX 8800
Страница 2: ......
Страница 41: ...Instruction Manual IM 106 880C OI Section 2 Installation April 2017 Section contents 17 ...
Страница 192: ...Section 7 Foundation Fieldbus April 2017 Instruction Manual IM 106 880C OI 16 6 About AMS and DeltaV Software ...
Страница 205: ...Alarm Relay Events April 2017 Instruction Manual IM 106 880C OI 17 8 Alarm Relay Events ...
Страница 251: ...Replace Tube Fittings April 2017 Instruction Manual IM 106 880C OI 22 4 Install Tube Fittings ...
Страница 259: ...Section 10 Replacement Parts April 2017 Instruction Manual IM 106 880C OI 23 2 O2 Cell and Heater Strut Assembly ...
Страница 261: ...Instruction Manual IM 106 880C OI Safety Data April 2017 Appendix contents 233 ...
Страница 263: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 235 ...
Страница 267: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 239 ...
Страница 271: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 243 ...
Страница 273: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 245 ...
Страница 275: ...Instruction Manual Safety Data IM 106 880C OI April 2017 IMPORTANT Safety Instructions 247 ...
Страница 277: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 249 ...
Страница 279: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 251 ...
Страница 298: ...SPA with HART Alarm April 2017 Instruction Manual IM 106 880C OI 27 2 Configuration Calibratio n ...
Страница 300: ...Instruction Manual IM 106 880C OI Index April 2017 Manual index 27 4 ...
Страница 304: ...Index April 2017 Instruction Manual IM 106 880C OI 27 8 Manual index ...
Страница 305: ......