Instruction Manual
IM-106-880C, OI
OCX 8800 Removal and
Installation
April 2017
Install Hazardous Area OCX
8800
18
3
9.3.2
Install Hazardous Area OCX 8800
Observe the following cautions when installing the Hazardous Area OCX 8800 in a hot process
stack. If the process is shut down and cooled, the transmitter can be installed in the stack prior
to connecting the pneumatics and wiring.
Whenever a positive stack pressure exists at the installation site, be sure to connect all
pneumatic lines prior to installing the Hazardous Area OCX 8800 in the stack or
ductwork. Failure to connect the pneumatic lines can allow the flow of contaminants
into the Hazardous Area OCX 8800 ports.
1.
Bolt the Hazardous Area OCX 8800 to the stack and install insulation. Refer to
and make sure all test gas lines and electrical connections are complete.
2.
Connect the test gas lines and the instrument air lines to the electronics housing.
3.
Remove the electronics housing cover.
4.
Install customer power and signal conduits and wiring at the electronics housing.
5.
If used, connect external relay leads to the alarm output relay terminal block,
6.
Connect the O
2
and COe signal leads to the 4-20 mA signal output terminal block.
7.
Connect the line (L1 wire) to the L1 terminal, and the neutral (N wire) to the N terminal
on the AC power input terminal block.
8.
Connect the ground lead to the ground stud. Secure the connection with two nuts.
Attach a separate ground lead (G wire) from the ground stud to the G terminal on the
power input terminal block.
9.
Install the cover on the electronics housing.
10.
Restore power to the system. Allow OCX to reach normal operating temperature.
11.
Turn on the test gasses at the cylinders and open the instrument air supply valve.
9.3.3
OCX with Remote Electronics
Remove Sensor Housing
1.
Turn off power to the system.
2.
Shut off the test gasses at the cylinders and close the instrument air valve.
3.
Disconnect the calibration gas, reference air, eductor air, and dilution air lines from the
sensor housing,
Содержание Rosemount OCX 8800
Страница 2: ......
Страница 41: ...Instruction Manual IM 106 880C OI Section 2 Installation April 2017 Section contents 17 ...
Страница 192: ...Section 7 Foundation Fieldbus April 2017 Instruction Manual IM 106 880C OI 16 6 About AMS and DeltaV Software ...
Страница 205: ...Alarm Relay Events April 2017 Instruction Manual IM 106 880C OI 17 8 Alarm Relay Events ...
Страница 251: ...Replace Tube Fittings April 2017 Instruction Manual IM 106 880C OI 22 4 Install Tube Fittings ...
Страница 259: ...Section 10 Replacement Parts April 2017 Instruction Manual IM 106 880C OI 23 2 O2 Cell and Heater Strut Assembly ...
Страница 261: ...Instruction Manual IM 106 880C OI Safety Data April 2017 Appendix contents 233 ...
Страница 263: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 235 ...
Страница 267: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 239 ...
Страница 271: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 243 ...
Страница 273: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 245 ...
Страница 275: ...Instruction Manual Safety Data IM 106 880C OI April 2017 IMPORTANT Safety Instructions 247 ...
Страница 277: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 249 ...
Страница 279: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 251 ...
Страница 298: ...SPA with HART Alarm April 2017 Instruction Manual IM 106 880C OI 27 2 Configuration Calibratio n ...
Страница 300: ...Instruction Manual IM 106 880C OI Index April 2017 Manual index 27 4 ...
Страница 304: ...Index April 2017 Instruction Manual IM 106 880C OI 27 8 Manual index ...
Страница 305: ......