Instruction Manual
IM-106-880C, OI
Section 5:
Calibration
April 2017
Manual O2/COe Calibration using the Field Communicator -
Fieldbus
73
5.5.5
Manual O
2
/COe Calibration using the Field Communicator -
Fieldbus
To perform a manual O
2
/COe calibration using the 375 Field Communicator, use the following
procedure. If necessary, refer to
Section 6, Field Communicator
, for the Fieldbus menu tree.
NOTE
To select a menu item, either use the up and down arrow keys to scroll to the menu item and
press the right arrow key or use the number keypad to select the menu item number.
To return to a preceding menu, press the left arrow key.
1.
To calibrate from Fieldbus, the first step is to set the Transducer Block to Out of Service
Mode (OOS).
2.
To set the OOS mode, select Transducer then select Process, followed by select Out of
Service in the Target Mode.
3.
From Transducer screen select Methods.
4.
Select OCX Calibration from the Methods menu.
5.
From the OCX Calibration screen, select Calibrate O2 Sensor for O2 calibration, select
Calibrate Combustibles Sensor for COe calibration, or select Calibrate Both Sensors if
calibrate both sensors is desired.
6.
In the calibration screen, a Loop should be removed from automatic control warning
appears. Remove the device from any automatic control loops to avoid a potentially
dangerous operating condition and press OK.
7.
The Select Action screen should look like the following. From the Select Action screen,
select Update Display to refresh the calibration status, select Next Calibration Step to
continue calibration, select Abort Calibration to abort calibration or Exit to exit
calibration. Select one item from the list and press OK. The Select Action screen is
static and data will not be automatically refreshed.
•
Calibration Step = Idle
•
Step Time Remaining = 0 seconds
•
O2 Value = 0.40 %
•
Combustibles Value = 1000 ppm
–
1. Update Display
–
2. Next Calibration Step
–
3. Abort Calibration/Exit
Содержание Rosemount OCX 8800
Страница 2: ......
Страница 41: ...Instruction Manual IM 106 880C OI Section 2 Installation April 2017 Section contents 17 ...
Страница 192: ...Section 7 Foundation Fieldbus April 2017 Instruction Manual IM 106 880C OI 16 6 About AMS and DeltaV Software ...
Страница 205: ...Alarm Relay Events April 2017 Instruction Manual IM 106 880C OI 17 8 Alarm Relay Events ...
Страница 251: ...Replace Tube Fittings April 2017 Instruction Manual IM 106 880C OI 22 4 Install Tube Fittings ...
Страница 259: ...Section 10 Replacement Parts April 2017 Instruction Manual IM 106 880C OI 23 2 O2 Cell and Heater Strut Assembly ...
Страница 261: ...Instruction Manual IM 106 880C OI Safety Data April 2017 Appendix contents 233 ...
Страница 263: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 235 ...
Страница 267: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 239 ...
Страница 271: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 243 ...
Страница 273: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 245 ...
Страница 275: ...Instruction Manual Safety Data IM 106 880C OI April 2017 IMPORTANT Safety Instructions 247 ...
Страница 277: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 249 ...
Страница 279: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 251 ...
Страница 298: ...SPA with HART Alarm April 2017 Instruction Manual IM 106 880C OI 27 2 Configuration Calibratio n ...
Страница 300: ...Instruction Manual IM 106 880C OI Index April 2017 Manual index 27 4 ...
Страница 304: ...Index April 2017 Instruction Manual IM 106 880C OI 27 8 Manual index ...
Страница 305: ......