![Emerson Rosemount OCX 8800 Скачать руководство пользователя страница 73](http://html.mh-extra.com/html/emerson/rosemount-ocx-8800/rosemount-ocx-8800_instruction-manual_104777073.webp)
Section 3: Configuration and
Startup
April 2017
Instruction Manual
IM-106-880C, OI
48
Configuring the Calibration Solenoids with the
LOI
4.
From the O2 CAL SETUP/COE CAL SETUP, select Solenoids. Select Present to enable the
solenoids.
3.5.3
Configuring the Calibration Solenoids with the LOI
1.
Use the "Z" pattern to enter the LOI menu tree.
2.
From the SYSTEM menu, select Calib Setup.
3.
From the Calib Setup menu, select Use Solenoids. Select Yes to enable the solenoids.
3.6
Blowback Feature
The blowback feature blows instrument air back through the center of the internal filter and out
the sample tube of the probe. This removes built up dirt and particulate from the internal filter,
sample line and any optional in-situ filter on the end of the sample tube. The blowback feature is
normally used in systems that have heavy particulate in the process stream.
The blowback feature requires the optional blowback hardware to be properly installed external
to the Hazardous Area OCX 8800. See Section 2, Pneumatic Installation, for details.
A Hazardous Area OCX 8800 shipped from the factory must be configured before blowback can
be implemented. This same process must be performed any time a replacement card stack is
installed.
3.6.1
Configuring Blowback with the Field Communicator - HART
1.
Use the 375 Field Communicator or AMS software to access the HART menu.
2.
From the DETAILED SETUP menu, select INPUT/OUTPUT.
3.
From the INPUT/OUTPUT menu, select BLOWBACK.
4.
From the BLOWBACK, select BlBk Enabled. Select Yes to enable blowback. Also set the
following parameters:
•
BlBk Intrvl - Length of time between blowback events (60 minutes
recommended).
•
BlBk Period - Length of time blowback is activated (5 seconds recommended).
•
BlBk Purge Time - Length of time after blowback is complete before oxygem/
combustibles readings are considered valid (Set as required by the application).
5.
Manually initiate blowback from DIAG/SERVICE, then BLOW BACK, When select
BLOWBACK.
6.
Configuring Blowback with the Field Communicator - Fieldbus
3.6.2
Use the 375 Field Communicator or AMS software to access
the Fieldbus menu.
1.
From the TRANSDUCER block menu, select Alarm Relay/Blowback.
Содержание Rosemount OCX 8800
Страница 2: ......
Страница 41: ...Instruction Manual IM 106 880C OI Section 2 Installation April 2017 Section contents 17 ...
Страница 192: ...Section 7 Foundation Fieldbus April 2017 Instruction Manual IM 106 880C OI 16 6 About AMS and DeltaV Software ...
Страница 205: ...Alarm Relay Events April 2017 Instruction Manual IM 106 880C OI 17 8 Alarm Relay Events ...
Страница 251: ...Replace Tube Fittings April 2017 Instruction Manual IM 106 880C OI 22 4 Install Tube Fittings ...
Страница 259: ...Section 10 Replacement Parts April 2017 Instruction Manual IM 106 880C OI 23 2 O2 Cell and Heater Strut Assembly ...
Страница 261: ...Instruction Manual IM 106 880C OI Safety Data April 2017 Appendix contents 233 ...
Страница 263: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 235 ...
Страница 267: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 239 ...
Страница 271: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 243 ...
Страница 273: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 245 ...
Страница 275: ...Instruction Manual Safety Data IM 106 880C OI April 2017 IMPORTANT Safety Instructions 247 ...
Страница 277: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 249 ...
Страница 279: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 251 ...
Страница 298: ...SPA with HART Alarm April 2017 Instruction Manual IM 106 880C OI 27 2 Configuration Calibratio n ...
Страница 300: ...Instruction Manual IM 106 880C OI Index April 2017 Manual index 27 4 ...
Страница 304: ...Index April 2017 Instruction Manual IM 106 880C OI 27 8 Manual index ...
Страница 305: ......