Instruction Manual
IM-106-880C, OI
Repair Sensor
Housing
April 2017
Remove
Eductor
20
1
Disassemble O
2
Cell and Heater Strut Assembly
Do not attempt to replace the O
2
cell until all other possibilities for poor performance have been
considered. If cell replacement is needed, order the O
2
cell replacement kit (Refer to
Section 10,
Replacement Parts
).
The O
2
cell replacement kit contains an O
2
cell and flange assembly, gaskets, socket head cap
screws, and anti-seize compound. The items are carefully packaged to preserve precise surface
finishes.
Do not remove items from the package until they are ready to be used.
1.
Remove the four Allen cap screws (1,
) from the O
2
cell (2). Remove the O
2
cell. The cell flange has a notch that may be used to gently pry the flange away from
heater tube (3).
NOTE:
The pad on the end of contact/thermocouple assembly (4) will sometimes fuse to the O
2
cell (2).
2.
If the O
2
cell is fused to the contact pad, push the O
2
cell back into the heater tube
(against spring pressure) and quickly twist the O
2
cell. The cell and contact pad should
separate. If the contact pad stays fused to the cell, a new contact/thermocouple
assembly (4) must be installed.
3.
Remove and discard gasket (5). Clean the mating surface of heater tube (3). Remove
burrs and raised surfaces with a block of wood and crocus cloth.
Use care when handling contact and thermocouple assembly. The ceramic rod in this
assembly is fragile.
4.
Remove screws (6), lockwashers (7), return wire (8), and heater strut assembly (9).
5.
If replacing contact and thermocouple assembly (4), use a pencil to mark location of
spring clip (10) before removing. Squeeze tabs on spring clip to remove. Retain spring
clip and spring (11); replace if damaged.
6.
While carefully handling new contact and thermocouple assembly (4) lay old assembly
next to new one. Transfer match marks to new assembly.
7.
Carefully guide new contact and thermocouple assembly (4) through strut bracket
(12), spring (11), and spring clip (10) until spring clip reaches pencil mark.
Содержание Rosemount OCX 8800
Страница 2: ......
Страница 41: ...Instruction Manual IM 106 880C OI Section 2 Installation April 2017 Section contents 17 ...
Страница 192: ...Section 7 Foundation Fieldbus April 2017 Instruction Manual IM 106 880C OI 16 6 About AMS and DeltaV Software ...
Страница 205: ...Alarm Relay Events April 2017 Instruction Manual IM 106 880C OI 17 8 Alarm Relay Events ...
Страница 251: ...Replace Tube Fittings April 2017 Instruction Manual IM 106 880C OI 22 4 Install Tube Fittings ...
Страница 259: ...Section 10 Replacement Parts April 2017 Instruction Manual IM 106 880C OI 23 2 O2 Cell and Heater Strut Assembly ...
Страница 261: ...Instruction Manual IM 106 880C OI Safety Data April 2017 Appendix contents 233 ...
Страница 263: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 235 ...
Страница 267: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 239 ...
Страница 271: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 243 ...
Страница 273: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 245 ...
Страница 275: ...Instruction Manual Safety Data IM 106 880C OI April 2017 IMPORTANT Safety Instructions 247 ...
Страница 277: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 249 ...
Страница 279: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 251 ...
Страница 298: ...SPA with HART Alarm April 2017 Instruction Manual IM 106 880C OI 27 2 Configuration Calibratio n ...
Страница 300: ...Instruction Manual IM 106 880C OI Index April 2017 Manual index 27 4 ...
Страница 304: ...Index April 2017 Instruction Manual IM 106 880C OI 27 8 Manual index ...
Страница 305: ......