Instruction Manual
IM-106-880C, OI
Section 2:
Installation
April 2017
Electrical
Connections
25
To maintain explosion-proof protection of the OCX88C in hazardous areas, the sample
entering the sensor housing must not exceed 195
C (383
F).
NOTE:
To maintain proper earth grounding, ensure a positive connection exists between the sensor
housing, the electronics housing, and earth. The connecting ground wire must be 14 AWG
minimum. Refer to
NOTE
Line voltage, signal, and relay wiring must be rated for at least 105 ºC (221 ºF).
2.4.1
Electrical Connections
Electrical connections, power and communications are made to the electronic enclosure. The
connections are made through two 3/4 NPT ports in the enclosure using fittings and cables
provided by the customer. Cable installation must meet NEC, IEC and/or other applicable
national or local codes for Class I, Zone 1, IIB +H2 T3/T6 permanently mounted equipment.
2.4.2
Connect Line Voltage
The Hazardous Area OCX 8800 operates on 100 to 240 VAC line voltage at 50 to 60 Hz. The
power supply requires no setup. Connect the line (L wire) to the L terminal, and the neutral (N
wire) to the N terminal on the AC power input terminal block in the electronics housing.
Connect the ground (G wire) to the ground stud in the electronics housing as shown in
2.4.3
Connect Output Signals
The Hazardous Area OCX 8800 may be provided with either two 4-20 mA signals with HART on
the O2 signal or a single Foundation Fieldbus signal. Connect the output terminals in the
electronics housing as shown in
. Use individual shielded twisted wire pairs. Terminate
the shield at the electronics housing.
2.4.4
O
2
4-20 mA Signal
One 4-20 mA signal represents the O
2
value. Superimposed on the O
2
signal is the HART
information accessible through a Model 375 Handheld Communicator or AMS Device Manager
software. The O
2
signal is at the AOUT 1 terminals.
Содержание Rosemount OCX 8800
Страница 2: ......
Страница 41: ...Instruction Manual IM 106 880C OI Section 2 Installation April 2017 Section contents 17 ...
Страница 192: ...Section 7 Foundation Fieldbus April 2017 Instruction Manual IM 106 880C OI 16 6 About AMS and DeltaV Software ...
Страница 205: ...Alarm Relay Events April 2017 Instruction Manual IM 106 880C OI 17 8 Alarm Relay Events ...
Страница 251: ...Replace Tube Fittings April 2017 Instruction Manual IM 106 880C OI 22 4 Install Tube Fittings ...
Страница 259: ...Section 10 Replacement Parts April 2017 Instruction Manual IM 106 880C OI 23 2 O2 Cell and Heater Strut Assembly ...
Страница 261: ...Instruction Manual IM 106 880C OI Safety Data April 2017 Appendix contents 233 ...
Страница 263: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 235 ...
Страница 267: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 239 ...
Страница 271: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 243 ...
Страница 273: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 245 ...
Страница 275: ...Instruction Manual Safety Data IM 106 880C OI April 2017 IMPORTANT Safety Instructions 247 ...
Страница 277: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 249 ...
Страница 279: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 251 ...
Страница 298: ...SPA with HART Alarm April 2017 Instruction Manual IM 106 880C OI 27 2 Configuration Calibratio n ...
Страница 300: ...Instruction Manual IM 106 880C OI Index April 2017 Manual index 27 4 ...
Страница 304: ...Index April 2017 Instruction Manual IM 106 880C OI 27 8 Manual index ...
Страница 305: ......