Section 5:
Calibration
April 2017
Instruction Manual
IM-106-880C, OI
68
Manual O2 Calibration using the Field Communicator -
HART
5.5.2
Manual O
2
Calibration using the Field Communicator -
HART
To perform a manual O
2
calibration using the HART communicator or AMS, use the following
procedure. If necessary, refer to Section 4, Using HART Communications, for the HART menu
tree.
NOTE:
To select a menu item, either use the up and down arrow keys to scroll to the menu item and
press the right arrow key or use the number keypad to select the menu item number.
To return to a preceding menu, press the left arrow key.
1.
Select DIAG/SERVICE from DEVICE SETUP menu.
2.
Select CALIBRATION from the DIAG/SERVICE menu.
3.
Select CAL CONTROL from the CALIBRATION menu.
4.
Select CAL METHODS from the CAL CONTROL menu.
5.
From the CAL METHODS menu, select the type of calibration desired: O2 Calibration.
6.
In the first Calibration screen, a Loop should be removed from automatic control
warning appears. Remove the Hazardous Area OCX 8800 from any automatic control
loops to avoid a potentially dangerous operating condition and press OK.
7.
The Calibration screen should look like the following. Press OK to continue.
•
OCX: TAG NAME
•
STATUS: Idle
•
TIME REMAIN: 0s
•
O2: 0.4 %, 85.95mV
•
OK/NEXT to Select
•
ABORT/CANCEL to Exit
8.
From the SELECT ACTION screen, select START/NEXT CALSTEP to continue calibration,
select ABORT CAL to abort calibration or EXIT CAL to exit calibration. Select one item
from the list and press ENTER.
•
OCX: TAG NAME
•
SELECT ACTION
–
1. START/NEXT CALSTEP
–
2. ABORT CAL
–
3. EXIT CAL
9.
When the Calibration Status is at the AppO2Low step, switch on O2 Low Gas. Verify the
O2 concentration measured matches the O2 LOW GAS parameter in the Setup. Press
OK when ready.
10.
Select Start/Next Cal Step to start applying the O2 Low Gas. The time to apply the test
gas is specified by the Gas Time.
Содержание Rosemount OCX 8800
Страница 2: ......
Страница 41: ...Instruction Manual IM 106 880C OI Section 2 Installation April 2017 Section contents 17 ...
Страница 192: ...Section 7 Foundation Fieldbus April 2017 Instruction Manual IM 106 880C OI 16 6 About AMS and DeltaV Software ...
Страница 205: ...Alarm Relay Events April 2017 Instruction Manual IM 106 880C OI 17 8 Alarm Relay Events ...
Страница 251: ...Replace Tube Fittings April 2017 Instruction Manual IM 106 880C OI 22 4 Install Tube Fittings ...
Страница 259: ...Section 10 Replacement Parts April 2017 Instruction Manual IM 106 880C OI 23 2 O2 Cell and Heater Strut Assembly ...
Страница 261: ...Instruction Manual IM 106 880C OI Safety Data April 2017 Appendix contents 233 ...
Страница 263: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 235 ...
Страница 267: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 239 ...
Страница 271: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 243 ...
Страница 273: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 245 ...
Страница 275: ...Instruction Manual Safety Data IM 106 880C OI April 2017 IMPORTANT Safety Instructions 247 ...
Страница 277: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 249 ...
Страница 279: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 251 ...
Страница 298: ...SPA with HART Alarm April 2017 Instruction Manual IM 106 880C OI 27 2 Configuration Calibratio n ...
Страница 300: ...Instruction Manual IM 106 880C OI Index April 2017 Manual index 27 4 ...
Страница 304: ...Index April 2017 Instruction Manual IM 106 880C OI 27 8 Manual index ...
Страница 305: ......