Figure 5-19: Select Gas (NO)
5. Use the middle buttons ( and ) to scroll up or down.
Select
RUN SPAN
.
Press the
CAL
button to calibrate, or the
VER
button to verify.
Figure 5-20: Action (SPAN)
6. Using the
LEFT, RIGHT, INCR
(increase) and
DECR
(decrease) buttons enter the
cylinder concentration (e.g. 20 ppm)
7.
Figure 5-21
displays as the calibration progresses.
Quick Start Guide
Calibrating the analyzer
D-7010-0055
August 2018
Quick Start Guide
69