WARNING
HAZARDOUS GAS
In accordance with IEC 60079-0: 2011 Clause 30.
The product stream that the analyzer is examining may be hazardous even at low concentrations. Special care should therefore be
taken to ensure that the sample gas return port either returns the sample gas to the product stream or discharges the sample gas
to a location that will not cause a hazard.
WARNING
HIGH PRESSURE GAS AND AIR
The calibration gas supply and compressed air supply operate at a pressure that can cause injury, e.g., damage to eyes and skin
punctures from debris blown by the high pressure gas or compressed air. Always lock off or tag out the calibration gas supply and
compressed air supply when shutting down the analyzer.
WARNING
HEAVY INSTRUMENT
The analyzer weighs 55 kg (121 lb.) and is designed to be wall mounted.
It must be moved in accordance with local safety regulations, however, Emerson recommend that a minimum of two people using
suitable tools for transportation and lifting are employed.
Use suitable fasteners for the weight of the unit. Make sure the wall the unit is mounted on is solid, stable and of a suitable
material to support the weight.
Failure to observe this warning and or follow safety instruction could cause an explosion or potentially hazardous situation, which
if not avoided, could result in death or serious injury.
WARNING
EXPLOSION HAZARD / ELECTRIC SHOCK HAZARD
In accordance with IEC 60079-0: 2011 Clause 30.
Only trained, qualified personnel may install and connect power and signal cables. The installation/connection must be in
accordance with all legislative requirements and applicable standards.
Failure to follow may cause warranty invalidation, property damage and/or personal injury or death.
Installation of the analyzer is subject to qualified personnel only, familiar with the resulting potential risks.
Analyzers providing screw terminals for electrical connections may require working near live part. Failure to observe this warning
and or follow safety instruction c could cause an explosion or potentially hazardous situation, which if not avoided, could result in
death or serious injury.
WARNING
HIGH PRESSURE HAZARD
In accordance with IEC 60079-0: 2011 Clause 30.
The maximum inlet purge gas pressure at the inlet valve must not exceed 690 kPa (6.9 bar). Higher pressure may damage the
analyzer enclosure in case of failure of the inlet valve.
Failure to observe this warning could cause a potentially hazardous situation which if not avoided, could result in death or serious
injury.
6
Содержание Rosemount CT5100 Ex
Страница 1: ...Quick Start Guide D 7010 0055 Rev C August 2018 Rosemount CT5100 Ex Continuous Gas Analyzer ...
Страница 12: ...Contents Quick Start Guide August 2018 D 7010 0055 12 Rosemount CT5100 Ex ...
Страница 34: ...Install Quick Start Guide August 2018 D 7010 0055 34 Rosemount CT5100 Ex ...
Страница 44: ...Operating the analyzer Quick Start Guide August 2018 D 7010 0055 44 Rosemount CT5100 Ex ...
Страница 80: ......
Страница 81: ......
Страница 82: ......
Страница 83: ......
Страница 84: ......
Страница 85: ......