Figure 2-7: Signal Cable Outputs
2.3.5
Power input cables and circuit breakers
WARNING
FLAMMABLE/EXPLOSIVE GASES
The purge / pressurization cycle expels any flammable or explosive gases from the
electrical compartment and
MUST
be completed before applying electrical power to the
rest of the analyzer.
Failure to observe this warning could cause an explosion or potentially hazardous
situation, which if not avoided, may cause death, personal injury, and/or damage to
persons and/or property.
The power input cable / purge power cable circuit breakers can be tested as follows:
•
Set the Purge circuit breaker to
ON
. The purge / pressurization process will start
ONLY
when the K1 LED on the purge control panel comes on indicating the purge process is
complete can you proceed to the next step. Refer to
Figure 3-4
.
Quick Start Guide
Install
D-7010-0055
August 2018
Quick Start Guide
31