absorption lines of the gases. The light from each laser is routed through an optical path to
the analysis cell, which provides measurement of low concentrations of the subject gases.
An external sample handling system conditions the sample gas and draws it through the
analysis cell. The light exits the multi-pass analysis cell and is directed to a receiver in the
analyzer. The variation in the intensity of light in the vicinity of the absorption lines is
measured, and the concentration is determined using a comprehensive spectral fitting
routine.
There is no sample conditioning provided within the analyzer; the sampled gas must be
brought within the parameters shown in the
Installation, Operations, and Maintenance
Manual
,
Detailed system specifications
in the before entering the analyzer. Detailed
characteristics of the analyzer are also given in the
Installation, Operations, and
Maintenance Manual
,
Detailed system specifications
.
Table 2-1: Main Items of the Rosemount CT5100(Ex) Installation
Item
Name or
description
Supplied by
Part number
Quantity
Notes
1
Rosemount
CT5100(Ex)
Emerson
Rosemount
CT5100(Ex)
1
2
Rosemount
CT5100(Ex)
software package,
version 5.7.13
Emerson
N/A.Software is
embedded in the
system.
1
Version described
in manual
3
Gas handling
system
Customer
(optionally by
Emerson)
Customer Choice
or Emerson
1
4
Heated gas sample
line hose
Customer
Customer choice
1
5
Exhaust line hose
(for sample gas)
Customer
Customer choice
1
6
Reference gas
cylinders
(instrument gas)
for calibration
purposes
Customer
Customer choice
Dependent upon
number of gases
being measured
7
Pressure regulator Customer
Customer choice
1 per gas cylinder Required for
calibration
8
Pneumatic T-piece Customer
Customer choice
1
Required for
calibration
9
Excess flow line
Customer
Customer choice
1
Required for
calibration
10
Power cables to
Rosemount
CT5100(Ex)
Customer
Customer choice
1
11
Cables from
Rosemount
CT5100(Ex) to
control center
Customer
Customer choice
1
Install
Quick Start Guide
August 2018
D-7010-0055
24
Rosemount
™
CT5100 Ex
Содержание Rosemount CT5100 Ex
Страница 1: ...Quick Start Guide D 7010 0055 Rev C August 2018 Rosemount CT5100 Ex Continuous Gas Analyzer ...
Страница 12: ...Contents Quick Start Guide August 2018 D 7010 0055 12 Rosemount CT5100 Ex ...
Страница 34: ...Install Quick Start Guide August 2018 D 7010 0055 34 Rosemount CT5100 Ex ...
Страница 44: ...Operating the analyzer Quick Start Guide August 2018 D 7010 0055 44 Rosemount CT5100 Ex ...
Страница 80: ......
Страница 81: ......
Страница 82: ......
Страница 83: ......
Страница 84: ......
Страница 85: ......