Figure 5-21: Calibration
8. Press
OK
to confirm.
Once complete the program will automatically open
Span Factor For NO
(
Figure
5-22
).
Figure 5-22: Span Input
Note that if an excessively high or low span factor is calculated (<0.7 or > 1.3), for
example if the calibration gas is a different concentration from that set above, it will
be rejected and the previous span factor will be displayed and will continue to be
used.
9. Press
OK
. The
Result
screen(
Figure 5-23
) displays.
Calibrating the analyzer
Quick Start Guide
August 2018
D-7010-0055
70
Rosemount
™
CT5100 Ex
Содержание Rosemount CT5100 Ex