Figure 5-31: Calibr param (AUTO PERIOD)
5. To change the time and the units (days, hours, min and sec) between calibrations:
Use the
LEFT, RIGHT, INCR
(increase)
DECR
(decrease) buttons to set the number and
time interval.
Press the
OK
button.
Figure 5-32: (Calibration) Period
6. To adjust the length of time for the span gas to flush the cell:
Use the middle buttons ( and ) to scroll up or down select
PURGE TIME
.
Press the
SEL
button.
Quick Start Guide
Calibrating the analyzer
D-7010-0055
August 2018
Quick Start Guide
75