NOTICE
The item letters in this procedure can be found in
Procedure
1. Place the flame simulator in a safe area, not exceeding 104 °F (40 °C).
2. Release the locking screw.
3. Unscrew the battery back cover (D) counter-clockwise.
4. Unscrew the locking disc (C) clockwise.
5. Pull out the battery from the flame simulator.
6. Connect the battery to the charger.
7. Charge for a maximum of two to three hours.
8. Disconnect the charger.
9. Insert the battery in the flame simulator.
10. Screw on the locking disc (C).
11. Screw on the back cover (D).
12. Lock the back cover with the locking screw.
D.1.6
Replace the battery
NOTICE
The item letters shown in this procedure can be found in
.
Procedure
1. Place the flame simulator on a table in a safe area, not exceeding 104 °F (40 °C).
2. Release the locking screw.
3. Unscrew the battery back cover (D) counter-clockwise.
4. Unscrew the locking disk (C) clockwise.
5. Pull out the battery from the flame simulator.
6. Insert the new battery pack in the simulator housing.
Use only Rosemount battery pack, PN 00975-9000-0012.
7. Screw on the locking disc (C).
8. Screw on the back cover (D).
9. Lock the back cover with the locking screw.
NOTICE
For more information, refer to the
Rosemount FS-HR-975 Flame Simulator Manual
Accessories
Reference Manual
June 2021
00809-0100-4976
70
Emerson.com/Rosemount
Legacy
Содержание Rosemount 975HR
Страница 28: ...Introduction Reference Manual June 2021 00809 0100 4976 28 Emerson com Rosemount L e g a c y ...
Страница 42: ...Installing the detector Reference Manual June 2021 00809 0100 4976 42 Emerson com Rosemount L e g a c y ...
Страница 66: ...RS 485 communication network Reference Manual June 2021 00809 0100 4976 66 Emerson com Rosemount L e g a c y ...
Страница 78: ...SIL 2 features Reference Manual June 2021 00809 0100 4976 78 Emerson com Rosemount L e g a c y ...