Table A-2: Electrical Specifications
(continued)
Operating voltage
Status
All outputs
Without 0-20 mA
Alarm when heater on
3.24 W
2.88 W
Maximum current
(max. 24 Vdc)
Normal
70 mA
65 mA
Normal when heater
on
95 mA
90 mA
Alarm
110 mA
95 mA
Alarm when heater on
135 mA
120 mA
Power consumption
(max. 18-32 Vdc)
Normal
1.95 W
1.85 W
Normal when heater
on
2.56 W
2.45 W
Alarm
3.04 W
2.56 W
Alarm when heater on
3.68 W
3.2 W
Maximum current (18 -
32 Vdc)
Normal
90 mA
85 mA
Normal when heater
on
105 mA
100 mA
Alarm
130 mA
115 mA
Alarm when heater on
160 mA
145 mA
Electrical input protection: The input circuit is protected against voltage-reversed polarity,
voltage transients, surges, and spikes according to MIL-STD-1275B.
A.3
Outputs
Electrical
interface
There are five output configurations. These configurations must be defined
at the factory per the customer order and cannot be changed at the
customer facility.
See
General instructions for electrical wiring
diagrams for each option.
Unless otherwise specified, the default is output configuration 1A. The
wiring arrangement is identified on the detector by the part number (see
•
Output configuration 1A: power, RS-485, 0-20 mA (sink), fault relay
(N.C.), alarm relay (N.O.) (see
•
Output configuration 2A: power, RS-485, 0-20 mA (source) and HART
®
protocol, fault relay (N.O.), alarm relay (N.O., N.C.)
•
Output configuration 3A: power, RS-485, 0-20 mA (source) and HART
protocol, fault relay (N.O.), alarm relay (N.O., N.C.)
•
Output configuration 1R: power, RS-485, fault relay (N.C.), auxiliary relay
(N.O.), alarm relay (N.O.)
Specifications and reference data
Reference Manual
June 2021
00809-0100-4976
52
Emerson.com/Rosemount
Legacy
Содержание Rosemount 975HR
Страница 28: ...Introduction Reference Manual June 2021 00809 0100 4976 28 Emerson com Rosemount L e g a c y ...
Страница 42: ...Installing the detector Reference Manual June 2021 00809 0100 4976 42 Emerson com Rosemount L e g a c y ...
Страница 66: ...RS 485 communication network Reference Manual June 2021 00809 0100 4976 66 Emerson com Rosemount L e g a c y ...
Страница 78: ...SIL 2 features Reference Manual June 2021 00809 0100 4976 78 Emerson com Rosemount L e g a c y ...