9. Click
Save
to save the new profile.
10. Click
OK
to close the
GC Directory
window.
11. Select
Chromatograph
→
Connect...
to connect to the GC or click
.
The
Connect to GC
window displays. The newly created GC profile should be listed in
the table.
12. Locate the new GC profile and click the
Ethernet
button that is associated with it.
The
Login
window displays.
13. Enter a User Name and User Pin and click
OK
.
4.4.13
Wiring the discrete digital inputs and outputs
The backplane has five discrete outputs and five discrete inputs. Refer to the
MON2020 Software for Gas Chromatographs Reference Manual
to learn how to configure
the digital outputs.
Related information
Wire a ROC800 digital output (DO) module
Wire the discrete digital inputs
WARNING
Electric shock
Failure to observe this precaution may cause serious personal injury or death.
The equipment operates using mains voltage that is dangerous to life. Make sure that the
circuit breakers are set to
OFF
and tagged off before removing the top cover or opening
the front cover.
WARNING
Explosion
Failure to de-energize the analyzer may cause an explosion and severely injure personnel.
Before opening the analyzer, disconnect all electrical power and ensure that the area is
free of explosive gases.
Keep cover tight while circuits are live.
Use cables or wires suitable for the marked "T" ratings.
Cover joints must be cleaned before replacing the cover.
Conduit runs to the enclosure must have sealing fitting adjacent to enclosure.
To connect digital signal input lines to the gas chromatograph (GC):
Procedure
1. Disconnect power to the analyzer and allow the components to cool for at least five
minutes.
2. Open the electronics enclosure door and access the back plane.
Reference Manual
Installation and start-up
2-3-9000-744
June 2022
Rosemount 700XA
75
Содержание Rosemount 700XA
Страница 1: ...Reference Manual 2 3 9000 744 Rev L June 2022 Rosemount 700XA Gas Chromatograph ...
Страница 6: ...TxD TD or Sout Transmit data or signal out 6 ...
Страница 30: ...Overview Reference Manual June 2022 2 3 9000 744 30 Emerson com Rosemount ...
Страница 100: ...Installation and start up Reference Manual June 2022 2 3 9000 744 100 Emerson com Rosemount ...
Страница 117: ...Figure 5 10 LSIV exploded view Reference Manual Operation and maintenance 2 3 9000 744 June 2022 Rosemount 700XA 117 ...
Страница 182: ...Operation and maintenance Reference Manual June 2022 2 3 9000 744 182 Emerson com Rosemount ...
Страница 223: ...Figure A 19 Chromatogram menu Reference Manual Local operator interface LOI 2 3 9000 744 June 2022 Rosemount 700XA 223 ...
Страница 262: ...Local operator interface LOI Reference Manual June 2022 2 3 9000 744 262 Emerson com Rosemount ...
Страница 270: ...Micro flame photometric detector µFPD Reference Manual June 2022 2 3 9000 744 270 Emerson com Rosemount ...
Страница 278: ...Pre defined Modbus map files Reference Manual June 2022 2 3 9000 744 278 Emerson com Rosemount ...
Страница 280: ......
Страница 281: ......
Страница 282: ......
Страница 293: ......