![Emerson Rosemount 700XA Скачать руководство пользователя страница 113](http://html.mh-extra.com/html/emerson/rosemount-700xa/rosemount-700xa_reference-manual_103388113.webp)
Replace solenoids
You can replace the oven system solenoids and/or the stream switching solenoids.
WARNING
Explosion
Failure to de-energize the analyzer may cause an explosion and severely injure personnel.
Before opening the analyzer, disconnect all electrical power and ensure that the area is
free of explosive gases.
Keep cover tight while circuits are live.
Use cables or wires suitable for the marked "T" ratings.
Cover joints must be cleaned before replacing the cover.
Conduit runs to the enclosure must have sealing fitting adjacent to enclosure.
Procedure
1. Remove the thermal cover from the upper enclosure.
2. Loosen the ultem thumb screw and tilt the oven on its side to gain access to the
solenoids that are located on the underside of the ultem.
3. Loosen the screws holding the solenoid in place and remove the solenoid.
4. Place a small amount of silicone grease on the target device (pneumatic block, four-
way stream block, etc.) where the solenoid is to be placed.
The grease ensures a tight seal.
5. Tighten the screws to hold the solenoid in place.
6. Place the oven upright.
7. Tighten the ultem thumb screw.
Reference Manual
Operation and maintenance
2-3-9000-744
June 2022
Rosemount 700XA
113
Содержание Rosemount 700XA
Страница 1: ...Reference Manual 2 3 9000 744 Rev L June 2022 Rosemount 700XA Gas Chromatograph ...
Страница 6: ...TxD TD or Sout Transmit data or signal out 6 ...
Страница 30: ...Overview Reference Manual June 2022 2 3 9000 744 30 Emerson com Rosemount ...
Страница 100: ...Installation and start up Reference Manual June 2022 2 3 9000 744 100 Emerson com Rosemount ...
Страница 117: ...Figure 5 10 LSIV exploded view Reference Manual Operation and maintenance 2 3 9000 744 June 2022 Rosemount 700XA 117 ...
Страница 182: ...Operation and maintenance Reference Manual June 2022 2 3 9000 744 182 Emerson com Rosemount ...
Страница 223: ...Figure A 19 Chromatogram menu Reference Manual Local operator interface LOI 2 3 9000 744 June 2022 Rosemount 700XA 223 ...
Страница 262: ...Local operator interface LOI Reference Manual June 2022 2 3 9000 744 262 Emerson com Rosemount ...
Страница 270: ...Micro flame photometric detector µFPD Reference Manual June 2022 2 3 9000 744 270 Emerson com Rosemount ...
Страница 278: ...Pre defined Modbus map files Reference Manual June 2022 2 3 9000 744 278 Emerson com Rosemount ...
Страница 280: ......
Страница 281: ......
Страница 282: ......
Страница 293: ......