Figure 1-8: µFPD PMT
A. µFPD PMT in the upper enclosure
The detection system in the µFPD uses the reactions of sulfur components in a
hydrogen/air flame as a source for analytical detection. The source of the µFPD's signal is
derived from the light produced by an excited molecule created in the flame's
combustion, that is, a photochemical process called chemiluminescence. A thermocouple
is fitted to the flame cell to ensure that the flame is present. If the flame is not detected,
the electrometer shuts off the hydrogen to the flame cell. It then supplies a voltage to the
igniter, waits five seconds, and opens the hydrogen shut off valve. The electrometer will
make between one and five ignition attempts if necessary. You can select the number of
ignition attempts on the
Hardware
→
Detector
screen. If the electrometer does not
succeed in igniting, then the GC shuts off the hydrogen, triggers an alarm, and waits for
attention from the operator.
Related information
Micro flame photometric detector (µFPD)
1.4.4
Micro flame photometric detector (µFPD) electronics
module
The electronics module contains two chambers. The internal chamber contains the photo
multiplier tube (PMT) to insulate it from outside temperature changes. The external
chamber is a thermo-electric cooler (TEC) controlled chamber, which houses the internal
chamber along with the electronic board that generates high voltage power through the
PMT.
Reference Manual
Overview
2-3-9000-744
June 2022
Rosemount 700XA
21
Содержание Rosemount 700XA
Страница 1: ...Reference Manual 2 3 9000 744 Rev L June 2022 Rosemount 700XA Gas Chromatograph ...
Страница 6: ...TxD TD or Sout Transmit data or signal out 6 ...
Страница 30: ...Overview Reference Manual June 2022 2 3 9000 744 30 Emerson com Rosemount ...
Страница 100: ...Installation and start up Reference Manual June 2022 2 3 9000 744 100 Emerson com Rosemount ...
Страница 117: ...Figure 5 10 LSIV exploded view Reference Manual Operation and maintenance 2 3 9000 744 June 2022 Rosemount 700XA 117 ...
Страница 182: ...Operation and maintenance Reference Manual June 2022 2 3 9000 744 182 Emerson com Rosemount ...
Страница 223: ...Figure A 19 Chromatogram menu Reference Manual Local operator interface LOI 2 3 9000 744 June 2022 Rosemount 700XA 223 ...
Страница 262: ...Local operator interface LOI Reference Manual June 2022 2 3 9000 744 262 Emerson com Rosemount ...
Страница 270: ...Micro flame photometric detector µFPD Reference Manual June 2022 2 3 9000 744 270 Emerson com Rosemount ...
Страница 278: ...Pre defined Modbus map files Reference Manual June 2022 2 3 9000 744 278 Emerson com Rosemount ...
Страница 280: ......
Страница 281: ......
Страница 282: ......
Страница 293: ......