Table 5-1: LSIV parts
ID number
Part number
Part name
01
125-004-XA-EM-SS
Vaporizer-cylinder, stainless steel (SS) 316
02
001-104-1/16-SS
Connector/carrier, 1/16-in, SS 316
03
120-112-1/8-PEEK
Gasket,
⅛
-in, polyetheretherketone (PEEK)
04
125-00801/16-SS
Connector/capillary-column, 1/16-in, SS 316
Equipped with VESPEL-ferrule, 1/16-in
05
001-131-1/16-V
VESPEL-ferrule, 1/16-in bore 0.5 mm/0.19-in
Usable for capillary column fused silica 0.32
06
101-228-1/8-SS
Connector,
⅛
-in, SS 316
(required only if
using packed column
⅛
-in or stream splitter)
07
120-038
Anchor plate
08
125-118-XA-EM-PEEK
Thermal barrier adapter flange (PEEK)
09
130-020-SS
Head piece, SS 316
10
001-164-SS
Screw, M4 x 12 hex socket countersunk head, SS
304
11
001-166-SS
Screw, M5x16 hex socket head cap, SS 304
12
No part number
13
No part number
14
No part number
Valve top unit
(1)
Denotes standard equipment.
(2)
See part ID 05.
(3)
See part ID 04.
(4)
Only if required.
(5)
Emerson equipment.
Procedure
1. Halt sample flow and allow time for the LSIV to cool.
2. Remove the actuation portion of the valve by unscrewing the union coupling from
the heater section, which should remain attached to the gas chromatograph (GC).
Emerson has provided a union coupling wrench for this purpose.
This will expose the sample flow chamber and the old seals that ride the metering
rod , which should be treated with great care to prevent bending or scratching.
3. Pull the sample flow chamber assembly off the metering rod. Remove the two seals.
To do this, you may need to push from the opposite side with a rod smaller than
0.175 in (4.4 mm).
4. Place new seals on the sample flow chamber assembly. Press the chamber and seals
back over the metering rod.
5. Place the actuation section in position on the heater section’s headpiece.
6. Use the union coupling spacer to ensure that the union coupling is properly aligned
with the heater section’s headpiece.
Operation and maintenance
Reference Manual
June 2022
2-3-9000-744
118
Emerson.com/Rosemount
Содержание Rosemount 700XA
Страница 1: ...Reference Manual 2 3 9000 744 Rev L June 2022 Rosemount 700XA Gas Chromatograph ...
Страница 6: ...TxD TD or Sout Transmit data or signal out 6 ...
Страница 30: ...Overview Reference Manual June 2022 2 3 9000 744 30 Emerson com Rosemount ...
Страница 100: ...Installation and start up Reference Manual June 2022 2 3 9000 744 100 Emerson com Rosemount ...
Страница 117: ...Figure 5 10 LSIV exploded view Reference Manual Operation and maintenance 2 3 9000 744 June 2022 Rosemount 700XA 117 ...
Страница 182: ...Operation and maintenance Reference Manual June 2022 2 3 9000 744 182 Emerson com Rosemount ...
Страница 223: ...Figure A 19 Chromatogram menu Reference Manual Local operator interface LOI 2 3 9000 744 June 2022 Rosemount 700XA 223 ...
Страница 262: ...Local operator interface LOI Reference Manual June 2022 2 3 9000 744 262 Emerson com Rosemount ...
Страница 270: ...Micro flame photometric detector µFPD Reference Manual June 2022 2 3 9000 744 270 Emerson com Rosemount ...
Страница 278: ...Pre defined Modbus map files Reference Manual June 2022 2 3 9000 744 278 Emerson com Rosemount ...
Страница 280: ......
Страница 281: ......
Страница 282: ......
Страница 293: ......