![Emerson Rosemount 700XA Скачать руководство пользователя страница 156](http://html.mh-extra.com/html/emerson/rosemount-700xa/rosemount-700xa_reference-manual_103388156.webp)
Procedure
1. Consult the parameter list that was provided with the gas chromatograph (GC) to
learn the appropriate flow rate.
2. Attach a flow meter to each measurement vent output on the side of the GC that is
labeled
MVn
, where
n
is for each vent.
The flow should match the value displayed in the parameter list.
5.4.11
Access electrical components
Emerson designed the gas chromatograph (GC) to operate for long periods of time
without needing preventative or regularly scheduled maintenance.
The enclosure is explosion-proof, dust-proof, water-proof, and flame-proof.
WARNING
Burns
Some parts of the analyzer may be heated to 248 °F (120 °C).
To prevent burns, do not touch any of the hot parts. All parts of an analyzer are always hot
unless it has been switched off and allowed to cool down.
Before fitting, removing, or performing any maintenance on the analyzer, make sure
that it has been switched off and allowed to cool for at least two hours.
When handling the analyzer, always use suitable protective gloves.
These precautions are particularly important when working at heights.
If burned, seek medical treatment immediately.
Prior to opening the GC, use MON2020 to ensure that there are no configuration or
parameter errors.
Procedure
1. Ensure electrical power is disconnected from the GC and the environment is safe.
2. Allow the GC to cool.
3. Unscrew and remove the front panel.
Operation and maintenance
Reference Manual
June 2022
2-3-9000-744
156
Emerson.com/Rosemount
Содержание Rosemount 700XA
Страница 1: ...Reference Manual 2 3 9000 744 Rev L June 2022 Rosemount 700XA Gas Chromatograph ...
Страница 6: ...TxD TD or Sout Transmit data or signal out 6 ...
Страница 30: ...Overview Reference Manual June 2022 2 3 9000 744 30 Emerson com Rosemount ...
Страница 100: ...Installation and start up Reference Manual June 2022 2 3 9000 744 100 Emerson com Rosemount ...
Страница 117: ...Figure 5 10 LSIV exploded view Reference Manual Operation and maintenance 2 3 9000 744 June 2022 Rosemount 700XA 117 ...
Страница 182: ...Operation and maintenance Reference Manual June 2022 2 3 9000 744 182 Emerson com Rosemount ...
Страница 223: ...Figure A 19 Chromatogram menu Reference Manual Local operator interface LOI 2 3 9000 744 June 2022 Rosemount 700XA 223 ...
Страница 262: ...Local operator interface LOI Reference Manual June 2022 2 3 9000 744 262 Emerson com Rosemount ...
Страница 270: ...Micro flame photometric detector µFPD Reference Manual June 2022 2 3 9000 744 270 Emerson com Rosemount ...
Страница 278: ...Pre defined Modbus map files Reference Manual June 2022 2 3 9000 744 278 Emerson com Rosemount ...
Страница 280: ......
Страница 281: ......
Страница 282: ......
Страница 293: ......