![Emerson Rosemount 700XA Скачать руководство пользователя страница 62](http://html.mh-extra.com/html/emerson/rosemount-700xa/rosemount-700xa_reference-manual_103388062.webp)
Figure 4-10: Sample stream inlets and calibration gas inlet
A. Sample stream inlets
B. Calibration gas inlet
4. Connect sample gas stream(s) to the GC.
•
Use
⅛
-in stainless steel tubing, as appropriate, to connect sample gas.
•
Unless stated otherwise in the product documentation, ensure that the pressure
of the calibration and sample line is regulated at 15 psig (1.03 barg) to 20 psig
(1.38 barg).
Postrequisites
After all lines have been installed, proceed with leak-checking the carrier and sample lines.
4.4.4
Maximum effective distance by communication
protocol type
lists the maximum distance at which the indicated protocol can transmit data
without losing effectiveness. If you need longer runs, use a repeater or other type of
extender to maintain the protocol's efficiency.
Table 4-1: Maximum distance for each communication protocol
Communication protocol
Maximum distance
RS-232
50 ft (15 m)
RS-422/RS-485
4,000 ft (1,219 m)
Installation and start-up
Reference Manual
June 2022
2-3-9000-744
62
Emerson.com/Rosemount
Содержание Rosemount 700XA
Страница 1: ...Reference Manual 2 3 9000 744 Rev L June 2022 Rosemount 700XA Gas Chromatograph ...
Страница 6: ...TxD TD or Sout Transmit data or signal out 6 ...
Страница 30: ...Overview Reference Manual June 2022 2 3 9000 744 30 Emerson com Rosemount ...
Страница 100: ...Installation and start up Reference Manual June 2022 2 3 9000 744 100 Emerson com Rosemount ...
Страница 117: ...Figure 5 10 LSIV exploded view Reference Manual Operation and maintenance 2 3 9000 744 June 2022 Rosemount 700XA 117 ...
Страница 182: ...Operation and maintenance Reference Manual June 2022 2 3 9000 744 182 Emerson com Rosemount ...
Страница 223: ...Figure A 19 Chromatogram menu Reference Manual Local operator interface LOI 2 3 9000 744 June 2022 Rosemount 700XA 223 ...
Страница 262: ...Local operator interface LOI Reference Manual June 2022 2 3 9000 744 262 Emerson com Rosemount ...
Страница 270: ...Micro flame photometric detector µFPD Reference Manual June 2022 2 3 9000 744 270 Emerson com Rosemount ...
Страница 278: ...Pre defined Modbus map files Reference Manual June 2022 2 3 9000 744 278 Emerson com Rosemount ...
Страница 280: ......
Страница 281: ......
Страница 282: ......
Страница 293: ......