Figure 3-33: Acquisition Module cable and transducer wiring
A. Base Enclosure gasket
B. Acquisition cable
C. Acquisition wiring terminal blocks
D. Acquisition Module
NOTICE
Ensure that the transducers identified as belonging to end 1 are installed on end 1
(A1, B1, C1 or D1) of the meter holder and those identified as belonging to end 2
(A2, B2, C2 or D2) are installed on end 2 of the meter housing.
6. Remove the three Acquisition Module flat head screws and split lock washers, then
remove the Acquisition Module from the Base Enclosure.
7. Insert the new Acquisition Module into the Base Enclosure and secure with the three
split lock washers and flat head screws.
8. Reattach the terminal blocks onto the Acquisition Module (3 mm flat head screw
driver required) for the corresponding chord (3414 four path meters A1, A2, B1, B2
C1, C2, D1, and D2; 3412 two path meters A1, A2, B1, B2; or 3411 single path
meters A1 and A2). Ensure the transducer wires have good contact with the
terminal block and the terminal block screws are tight.
9. When you have completed attaching the Transducer wire terminal blocks and the
Acquisition cable terminal block to the Acquisition Module, check the Base
Enclosure o-ring and reinstall if necessary.
10. Reattach the Transmitter Electronics Enclosure to the Base Enclosure with the two
hex head bolts and lock washers. Tighten bolts with a 6mm Allen wrench.
11. Retighten or reattach the conduit to the Transmitter Electronics Enclosure using a
crescent wrench or channel lock pliers.
Meter repairs
Maintenance and Troubleshooting Manual
April 2022
00809-0600-3104
88
Emerson.com/Rosemount