transducer stalk, next to its serial number (see
), i.e. -01, -02, ... -13 after
"T-200". If adjustment is needed:
a) Use one hand to hold the crystal holder and the other hand to hold the
transformer housing (see
b) Turn the transformer housing clock wise by 90 degrees and slide it slowly to
align the tab to the proper position according to
.
c) Turn the transformer housing counter-clock wise by 90° to let the tap snap
into the correct slot. Ensure the tap is secured in the slot.
4. Hold the transducer capsule assembly vertically and apply a small amount of
Acoustic Coupling Fluid (P/N: 1-360-01-650) to the surface of Kapton tape.
5. Carefully spread the coupling fluid on the surface of Kapton tape using the tip of the
plastic bottle for the coupling fluid.
Note
Do not press the Kapton tape while spreading the coupling fluid.
6. Wait for 30 - 60 seconds to let the coupling fluid spread more evenly on the surface.
7. Insert the transducer capsule assembly into the transducer stalk while aligning the
anti-rotation pin on the capsule with the slot on the inner surface of the stalk.
Figure 3-15: Length verification of T-200 transducer capsule assembly
A. Transducer stalk
B. Indicator ring
C. Capsule assembly
8. Make sure that the indicator ring on the transducer capsule assembly is flush with
the end of the transducer stalk (see
) to ensure the capsule is set to the
correct length.
9. Screw the transducer retainer on the stalk and use a 1
⅛
" wrench to bottom out.
Meter repairs
Maintenance and Troubleshooting Manual
April 2022
00809-0600-3104
66