Table 2-2: Troubleshooting
(continued)
Error
Recommended action(s)
Configuration changed
•
One or more parameters have been modified in the meter's
configuration,
—
Collect an Audit log using MeterLink in order to see what
configuration parameters changed and when they changed.
—
Run the
Tools
→
Edit/Compare Configuration
utility and select
Write All
or
Write Checked
values to write the changes to the
meter.
—
Save the configuration file.
Configuration lost
•
The meter configuration has reset to default values and the meter is
not configured correctly to measure flow and the meter has
performed a Cold Start.
—
Unless the Cold Start occurred after upgrading firmware,
replace the CPU board.
—
If the cold start occurred after a firmware upgrade, fully re-
configure the meter from a previously saved configuration
using the
Tools
→
Edit/Compare Configuration
in MeterLink.
Electronics Temperature
is Out of Nominal Range
•
Temperature of the internal electronics is out of nominal operating
range (below -40 °C or above 100 °C) which could lead to a system
failure.
—
Attempt to warm or cool the meter electronics housing.
—
Collect a Maintenance log using MeterLink while the meter is
experiencing the issue and contact your service representative.
Maintenance and Troubleshooting Manual
Troubleshooting
00809-0600-3104
April 2022
Rosemount Gas Ultrasonic Flow Meters
21