Figure 3-1: J-Mount transducer assembly
A. Transducer cable (max. length 15 ft.)
B. Transducer cable nut and chordset
C. Mount
D. Transducer assembly, stalk and holder
Figure 3-2: J-Mount transducer disassembly
A. Transducer cable (max. length 15 ft.)
B. Transducer cable nut and chordset
C. Transducer holder
D. Transducer stalk (optional)
E. Transducer
F. Mount
3.2.2
T-Slot transducer removal without extractor tool
This procedure requires venting the line pressure prior to removing the transducer(s).
Before removing and installing transducer(s), connect to the meter using MeterLink and
collect and save a Maintenance Log.
Note
The following instructions are for transducer removal and installation without the use of
an extractor tool. It is recommended that one work on one transducer assembly at a time
to reduce the possibility of improper assembly with respect to transducer lengths and
location.
•
Transducers are always replaced in pairs.
Maintenance and Troubleshooting Manual
Meter repairs
00809-0600-3104
April 2022