15. Insert the transducer capsule into the transducer stalk while aligning the anti-
rotation pin on the capsule with the slot on the inner surface of the stalk.
16. Make sure that the indicator ring on the transducer capsule is flush with the end of
the transducer stalk to ensure the capsule is set to the correct length (see
17. Hand-tighten the transducer retainer onto the stalk.
18. Follow
T-200 transducer assembly installation
to install the transducer assembly in a
meter body.
3.7
Transducer cable removal and installation
Rosemount
™
3410 Series Ultrasonic Gas Flow Meters have blue transducer cables and a
transducer cable nut that threads directly into the back of the transducer holder. The
cables use plastic glands that are installed on the meter’s Base Enclosure.
Note
Make a note of the exiting cabling path layout to allow proper tie-wrap configuration later
in this procedure.
Figure 3-17: Rosemount 3410 Series Ultrasonic Gas Flow Meter transducer cables and
ports
A. Base enclosure transducer cable glands
B. Cable ties (two locations)
C. Rosemount 3410 Series Ultrasonic Meter transducer port
D. Transducer mount and transducer cable nut and chordset (P/N 1-360-01-310)
Meter repairs
Maintenance and Troubleshooting Manual
April 2022
00809-0600-3104
70