5. Use non-fiber paper to remove any residue of acoustic coupling fluid on the surface
of Kapton tape.
6. Record the serial number of the removed transducer capsule assembly and put it in
a safe place.
7. Repeat
if more transducer capsule assemblies are to be removed.
3.5.2
T-200 transducer capsule assembly installation
This procedure is for installing a transducer capsule assembly into a transducer assembly
while it is installed in a meter body.
Procedure
1. Ensure that the transducer stalk, retainer and capsule assembly are clean and free of
debris.
2. Record the serial number of the transducer capsule assembly to be installed and
make sure it is correct for the intended transducer assembly.
Table 3-3: T-200 transducer stalk assembly configuration and capsule setting
Stalk assembly configuration
Smart capsule setting
-01
1
-02
2
-03
3
-04
4
-05
5
-06
1
-07
2
-08
3
-09
4
-10
5
-11
6
-12
7
-13
8
Meter repairs
Maintenance and Troubleshooting Manual
April 2022
00809-0600-3104
64
Emerson.com/Rosemount