F.3
FC03 register area
Read Holding Register.
F.3.1
System register area
Table F-3. System Register Area
Register
Number
Register Name
Type
Unit
Description
Unit Register Area
Priority 5 question. Defines the unit for
values in the FC04 area.
100
Level Unit
WORD
Enum (m)
44 = ft
45 = m (default)
101
Level Rate Unit
WORD
Enum (m/h)
120 = m/h (default)
247 = ft/h
102
Temperature Element Unit
WORD
Enum (1/10°)
1 = 1/10 ° (default)
2 = 1/100 °
Resolution for the value, the
Temperature Unit below will be used
103
Temperature Unit
WORD
Enum (°C)
32 = °C (default)
33 = °F
104
Pressure Unit
WORD
Enum (barG)
11 = Pa
12 = kPa
262 = psi A
263 = bar A
518 = psi G
519 = bar G (default)
239 = mm H2O
Содержание Rosemount 2460
Страница 2: ......
Страница 4: ...ii Reference Manual 00809 0100 2460 Rev DE Title Page October 2018 Title Page ...
Страница 12: ...x Reference Manual 00809 0100 2460 Rev DE Contents October 2018 Contents ...
Страница 18: ...6 Reference Manual 00809 0100 2460 Rev DE Introduction October 2018 Introduction ...
Страница 60: ...48 Reference Manual 00809 0100 2460 Rev DE Installation October 2018 Installation ...
Страница 100: ...88 Reference Manual 00809 0100 2460 Rev DE Configuration October 2018 Configuration ...
Страница 182: ...170 Product Certifications Product Certifications October 2018 Reference Manual 00809 0100 2460 Rev DE ...
Страница 214: ...202 Modbus Standard Question Modbus Standard Question October 2018 Reference Manual 00809 0100 2460 Rev DE ...
Страница 230: ...218 Varec Device Configuration Varec Device Configuration October 2018 Reference Manual 00809 0100 2460 Rev DE ...
Страница 238: ...226 Reference Manual 00809 0100 2460 Rev DE Index October 2018 Index ...
Страница 239: ......