Calibration test
Disconnect
Board A
from
Board B
of
AP3 diagnostic POD
Insert
Board A
of
AP3 diagnostic POD
into Programming Module Interface (PMI)
connectors of the programmer.
Run calibration test of programmer in PG4UW (menu Programmer / Calibration test).
AP3 diagnostic POD for Calibration test
Technical specification
HARDWARE
Base unit, DACs
USB 2.0 high-speed compatible port, up to 480 Mb/s transfer rate
100Mbit LAN port
on-board intelligence: powerful processor (ARM9 400MHz) and FPGA based state machine
(basic clock 50 MHz plus PLL)
built-in mSATA SSD as internal buffer (128GB, upgradeable to higher capacity) (*)
three D/A converters for VCC1, VCC2, and VPP, controllable rise and fall time
VCC1, VCC2 range 0.8V..7V/1A (step 10mV)
VPP range 0V..25V/1A (step 25mV)
Temperature controlled fan
selftest capability
protection against surge and ESD on power supply input, USB and LAN port and pins of
Programming Module Interface (PMI)(IEC1000-4-2: 15kV air, 8kV contact)
banana jack for ESD wrist straps connection
Содержание BeeHive304
Страница 6: ...Elnec s r o 6 Introduction ...
Страница 10: ...Elnec s r o 10 Quick Start ...
Страница 13: ...Detailed description 13 Detailed description ...
Страница 14: ...Elnec s r o 14 BeeHive304 ...
Страница 29: ...BeeProg3 29 BeeProg3 ...
Страница 43: ...Setup 43 Setup ...
Страница 48: ...Elnec s r o 48 Step 9 Installation process will start Step 10 Click Finish button to finish setup ...
Страница 58: ...Elnec s r o 58 PG4UW software ...
Страница 125: ...PG4UWMC software 125 PG4UWMC software ...
Страница 138: ...Elnec s r o 138 Installation procedure customized ...
Страница 143: ...Common notes 143 Common notes ...
Страница 148: ...Elnec s r o 148 Troubleshooting and warranty ...
Страница 152: ...Elnec s r o Jana Bottu 5 SK 080 01 Presov Slovakia www elnec com ZLI 0330 ...