13
Leakage current tester 92-4A / 92-4D / 92-2E
Safety Regulations
The operating instructions have to be always available at the operational
place of the tester. Incomplete or illegible operating instructions have to
be replaced immediately. Of course, we are pleased to be of help to you
in this respect.
In addition to the operating instructions legal, generally valid and other
obligatory regulations for accident prevention and environmental protection
have to be observed.
2.5
Further notes
Attention:
During leakage current measurements the enclosure of the item
under test can be voltage-carrying. Therefore, it is necessary to
take action to prevent a touching of the item under test during the
measurement ( separate test room, barring off, warning signs ...)
Attention: Before a leakage current measurement is conducted make
sure that the protective earth conductor connection is in good order!!
Therefore, prior to the leakage current measurement a protective
earth continuity test must be carried out under all circumstances.
Attention:
The unit is rather heavy (see chap. 1 "Technical Main Data "). Therefore
it is necessary, to take suitable actions during installation and removing.