17
Leakage current tester 92-4A / 92-4D / 92-2E
Make ready
3.6
Connection of the interfaces
3.6.1 Connection of the CAN-interface (Option)
Plug in the CAN-BUS-interface cable into the 9pole D-SUB-jack X13.
The next user in the bus system must be connected to X14.
If the control shall be carried out via the CAN-interface then the parameter
of the interface must be set in the menu to "
Equipment settings"
CAN-Bus.
Note:
If the LC-Tester is the last unit at the CAN-bus system, then the bus has
to be assigned to X14. The CAN-bus must be supplied with 10 -12 V DC.
3.6.2 Connection of the RS 232-interface
Plug in the serial interface cable into the 9pole D-SUB-jack X15. For con-
nection to the PC a so-called "Zero modem-cable" has to be used.
If the control shall be carried out via the RS232-interface then the pa-
rameter of the "RS232-interface" must be set in the menu to "
Equipment
settings-RS232"
(chapt. 4.6.4).
Note:
If the leakage current tester is controlled by a PC via the RS 232-inter-
face, then the protocol printer has to be connected to the PC. In menu
"
Equipment settings
- Print" (Chapt. 4.6.5) the printer must be deactivated.
3.6.3 Connection of the printer
A protocol printer can be connected to the RS 232 interface. If a printer is
connected then the parameter "Printer setting" in the menu "
Equipment
settings-Printing"
(Chapt. 4.6.5) must be set.
Use a so-called "Zero modem-cable" and if necessary a Serial/Parallel
converter.