Leakage current tester 92-4A / 92-4D / 92-2E
MENU SYSTEM
4.3
Menu Manual test
In the menu „Manual Test“ every single basic test step can explicitly be
selected and executed.
The initial settings of the test are saved in the
selected parameter block. The menu serves principally for creating and
testing new test data (type sample test).
Action buttons:
A1 ... A6
- after actuating the rotary encoder S4 the selected basic test
step will be prepared. The measurement is started by operating the
green RUN-button. The quantity of the displayed test steps (A1, A2,
A3, ....) depends on the pre-set test parameters. The basic test steps
are described in enclosure A5.
Operation
- The mains supply voltage is connected to the output termi-
nals.
Menu buttons:
Single / Group
-
(for a manually performed test)
Group
- Activates a basic test step (A1 ... A6) by means of rotary
encoder S4. Press the green RUN-button. The selected basic
test step will be processed completely, this means there is no
stop before the individual partial test step.
Single
- Activates a basic test step (A1 ... A6) by means of rotary
encoder S4. Press the green RUN-button. The selected basic
test step will be processed. Every individual partial test step
has to be explicitly confirmed by pressing the RUN-button.
Change
- the menu for setting the test parameter is displayed,
see Chap. 4.5.
Load
- another parameter block can be loaded, see Chap. 4.5.3 "Load
Parameter Block". The next measurement is then carried out with the
initial settings of the parameter block.
Save
- the currently set measurements can be stored as new parameter
block in one of the 200 available storages. See Chap. 4.5.4 "Save
Parameter Block".
Illustr. 4.3: Test monitor- Manual Test
Manual