55
Leakage current tester 92-4A / 92-4D / 92-2E
Testing
6.3
Automatic leakage current test
Prerequisite:
- The item under test is connected correctly (Chap. 6.1).
• Go to the menu "Automatic Test".
The status display for an automatic test is displayed
Attention:
During leakage current measurements the enclosure of the item
under test can be voltage-carrying. Therefore, it is necessary to
take action to prevent a touching of the item under test during the
measurement ( separate test room, barring off, warning signs ...)
• If required select with "Selection" a new parameter block.
• Select with rotary encoder S4 "START".
• Confirm "START" by pressing rotary encoder S4.
• Start the leakage current measurement by pressing the green "RUN"-but-
ton S3/H3.
The leakage current measurement will be started.
Warming-up time:
If a warming-up time is set (Chap. 4.5.1 Basic parameter), the oper-
ating voltage will be applied for the pre-set time. The display shows
the warming-up time.
Illustr. 6.4: Example: Automatic leakage current test for EN60065
RUN