26
Chapter 6 Reference
Environmental
Conditions
Temperature
Operating: 0 °C - 35 °C (32 °F - 95 °F)
Transportation/Storage: -20 °C - 60 °C (-4 °F - 140 °F)
Humidity
Operating: 20% - 80% R.H. (no condensation)
Transportation/Storage: 10% - 80% R.H. (no condensation)
Pressure
Operating: 700 to 1,060 hPa
Transportation/Storage: 200 to 1,060 hPa
USB
Standard
USB Specification Revision 2.0
Port
Upstream port × 1, Downstream port × 2
Communication Speed
480 Mbps (high), 12 Mbps (full), 1.5 Mbps (low)
Supply current
Downstream: Max. 500mA/1 port
Classification of Equipment
Type of protection against electric shock : Class I
EMC class: EN60601-1-2:2007 Group 1 Class B
Classification of medical device (MDD 93/42/EEC) : Class I
Mode of operation: Continuous
IP Class: IPX0
Main default settings (factory settings)
CAL Switch
1-DICOM
Power Save
On
Input Selection
Auto
Logo
Off
Menu Position
Center
Language
English
Signal Selection*
Single Link
Orientation*
Portrait (HW)
* These functions cannot be initialized with the All Reset function (
page 16
). For information about setting these functions, refer
to“3-2. Setting the Input Signals” (
page 13
) and "3-1. Setting Orientation" (
page 12
).