![Edwards nRVi Скачать руководство пользователя страница 50](http://html1.mh-extra.com/html/edwards/nrvi/nrvi_instruction-manual_2366141050.webp)
4.12 Connection for multi-drop
Using multi-drop mode, a single computer system can communicate with more than one
pump.
To assign individual addresses and to see the main differences in multi-drop message
protocol, refer to
Operation with multi-drop on page 58
.
Figure 14 RS485 multi-drop connections
1. RS485 interface on control equipment or
PC
2. nRVi pump logic interface PC
1. RS485 interface on control equipment or
PC
2. nRVi pump logic interface PC
4.13 Leak test the system
WARNING:
Process gases which leak from the pipelines will be dangerous to people and there
may be a danger of explosion or fire if air leaks into the pipelines
Uncontrolled air leaks may lead to flammable mixtures in the process gas pipelines. We
will accept no liability or warranty claims resulting from flammable mixtures caused by
air leaks.
Leak test the pipelines according to all local and national regulations, but as a minimum
the pump exhaust and the bypass pipelines must be tested.
1.
Leak test the system.
2.
Seal any leaks found.
Page 46
A65901880_D - Installation
Содержание nRVi
Страница 1: ...Rotary Vane Pumps nRVi INSTRUCTION MANUAL A65901880_D Original instructions edwardsvacuum com...
Страница 6: ...This page has been intentionally left blank...
Страница 97: ...Page 93 A65901880_D Service...
Страница 98: ...This page has been intentionally left blank...
Страница 99: ...This page has been intentionally left blank...
Страница 100: ...edwardsvacuum com...