3. Technical data
3.1 Operating and storage conditions
Table 2
Operating and storage conditions
Parameter
Value
Units
Ambient operating temperature range
*
12 to + 40
°C
Ambient storage temperature range
- 30 to + 70
°C
Normal surface temperature of the pump
body
§
+ 50 to + 70
°C
Maximum operating humidity
90
% RH
Maximum operating altitude
‡
2000
m
Installation category (BS5295)
II
-
Pollution degree (EN61010)
2
-
* Under cold start conditions the smart drive accelerates the pump in a controlled way
and it may take up to 90 seconds to reach full speed.
§ At ultimate vacuum, with ambient temperature of + 20 °C.
‡ The pump can operate safely up to 3000 m, when operating above 2000 m contact us
for advice.
Note:
To comply with IEC/UL/CSA 61010 standards, the pump must be installed and used
indoors, and within the operating conditions specified.
3.2 Mechanical data
Table 3
Mechanical data
Parameter
nRV4i
nRV6i
nRV10i
nRV14i
Units
Dimensions
Refer to
mm
Maximum tilt angle
10
degrees
Nominal rotational
speed
1800
rpm
Standby rotational
speed
Factory default set to 83% of nominal speed
Customer can set standby speed between 67% and
100% of nominal full speed
% rpm
Page 16
A65901880_D - Technical data
Содержание nRVi
Страница 1: ...Rotary Vane Pumps nRVi INSTRUCTION MANUAL A65901880_D Original instructions edwardsvacuum com...
Страница 6: ...This page has been intentionally left blank...
Страница 97: ...Page 93 A65901880_D Service...
Страница 98: ...This page has been intentionally left blank...
Страница 99: ...This page has been intentionally left blank...
Страница 100: ...edwardsvacuum com...