![EDAN F15 Series Скачать руководство пользователя страница 79](http://html1.mh-extra.com/html/edan/f15-series/f15-series_service-manual_2357161079.webp)
F15 Series Fetal & Maternal Monitor Service Manual
Connecting to OB TraceVue System
- 68 -
Figure A2-5 Connecting DB9 and RJ45 interfaces of the OB TraceVue system
Setting
Set the Net Version of F15/F15Air to
PHILIPS
. (Refer to
5.8 Changing Net Version
)
Restart F15/F15Air.
Validation
1) Switch F15/F15Air to Demo mode.
2) Observe the surveillance window of the OB TraceVue system. If data from F15/F15Air is
transferred to the system, the connection succeeds. Otherwise check the connection or try on
another interface of the system.
NOTE:
The identity number of this interface is preset. The device No. of
F15/F15Air does not take effect in
this system.
Содержание F15 Series