F15 Series Fetal & Maternal Monitor Service Manual
Troubleshooting
- 36 -
Bad
connection
of
electrodes.
Check the connection.
TOCO&MECG Transducer
defective.
Replace
the
TOCO&MECG
Transducer.
Main
control
board
defective.
Replace the main control board.
7.11 SpO2 Monitoring Failures
Phenomenon
Possible Cause
Solution
No
SpO
2
waveform
or
numeric value.
SpO
2
transducer defective.
Replace the SpO
2
transducer.
Bad connection of SpO
2
cable.
Check the connection.
iM X12
parameter board
defective.
Replace the
iM X12
parameter
board
.
Main
control
board
defective.
Replace the main control board.
7.12 NIBP Monitoring Failures
Phenomenon
Possible Cause
Solution
The cuff fails to be inflated.
Bad connection of the cuff,
the air hose and the NIBP
socket.
Check the connection from outside.
Bad connection of the iM
X12 board to the NIBP
socket.
Check the connection from inside.
Cuff
or
extension
tube
defective.
Replace the damaged part(s).
Air pump, air valve leak.
Replace pump valve module.
iM X12 parameter board
defective.
Replace the iM X12 parameter board.
Main
control
board
defective.
Replace the main control board.
No NIBP numeric values.
iM X12 parameter board
defective.
Replace the iM X12 parameter board.
Air pump, air valve leak.
Replace pump valve module.
Main
control
board
defective.
Replace the main control board.