![EDAN F15 Series Скачать руководство пользователя страница 23](http://html1.mh-extra.com/html/edan/f15-series/f15-series_service-manual_2357161023.webp)
F15 Series Fetal & Maternal Monitor Service Manual System Configuration
- 12 -
Chapter 4 System Configuration
The end users can not change the system configurations of the monitor. As a service engineer,
you need to change these configurations for them after the monitor is installed and checked
properly.
NOTE:
Restart the monitor after changing the settings.
4.1 Opening System Setup Menu
The system configurations of the monitor are to be changed in the system setup menu. To open
this menu,
1 Select the setup key
on the main interface.
2 Select System.
3 Input the password 9999 on the soft keyboard.
4 Select Enter on the soft keyboard.
4.2 Entering Demo Mode
The monitor works in real-time monitoring mode when monitoring a patient. If you want to show
the traces and parameters for a demonstration, you need to enter the Demo mode.
1 Open the System Setup menu.
2 Select
Demo
.
3 Select
OK
.
4.3 Reloading Default Settings
The default settings are predefined in the factory. Users can change the monitor setup, but this
default configuration stays the same. To reload this default,
1 Open the System Setup menu.
2 Select
Default
.
3 Select
OK
.