Modules’ Malfunction Verification
- 45 -
8.5 Verifying Malfunction of EDAN Transducer Socket Board
(wired model only)
To verify the EDAN transducer socket board,
1)
Open the main unit using the procedures described in section 9.3.
2)
Switch on the monitor. (Powered by AC or battery.)
3)
Measure the voltage to earth of the pins listed below, using a multimeter:
Item
Reference Pin
Test Pin
Reference result
○
1
TP24(GND)
TP23
4.75~5.25V
○
2
TP24(GND)
TP27
11.4~12.6V
○
3
TP25(FGND)
TP26
5.23-5.78V
○
4
TP25(FGND)
TP28
4.75-5.25V
EDAN Transducer Socket Board
4)
If any one of the results exceeds the reference range, the EDAN transducer socket board
defection is confirmed. Replacement of the board is recommended.
○