F15 Series Fetal & Maternal Monitor Service Manual System Configuration
- 16 -
192.168.1.X.
14
Select
OK
in the Network Setup menu.
15
Select
OK
in the System Setup menu and then restart the monitor.
4.11 Changing Trace Width
In order to differentiate the traces clearly, you may assign different widths for different traces.
1 Open the System Setup menu.
2 Select
Recorder Configuration
.
3 Select
FHR1
.
4 Select a number from
1
,
2
,
3
and
4
. The width of the trace increases by degrees from
1
to
4
.
5 Set width of FHR2, FHR3, AFM, TOCO, HR and SpO
2
traces in the same way.
6 Select
OK
in the Recorder Configuration menu.
7 Select
OK
in the System Setup menu.
4.12 Changing Paper Style
The paper style determines the range of FHR trace background pane bar. Observe the recorder
paper that the user has ordered, 30 ~ 240 is USA style; 50 ~ 210 is International style.
1 Open the System Setup menu.
2 Select
Recorder Configuration
.
3 Select the pane next to
Paper Style
.
4 Select the required style from
USA
and
International
.
5 Select
OK
in the Recorder Configuration menu.
6 Select
OK
in the System Setup menu.
CAUTION
To avoid inconsistency between FHR traces on the recorder paper and those on the
screen, the paper style must comply with the style of the recorder paper being used.
4.13 Changing the Language
The software supports several languages.
1 Open the System Setup menu.
2 Select
Language
.
3 Select the required language.
4 Select
OK
.
Содержание F15 Series
Страница 1: ......
Страница 15: ...F15 Series Fetal Maternal Monitor Service Manual Safety Guidance 4 42 DO NOT STEP ON ...
Страница 51: ...F15 Series Fetal Maternal Monitor Service Manual Modules Malfunction Verification 40 Main control board TP61 ...
Страница 75: ...F15 Series Fetal Maternal Monitor Service Manual Periodic Tests 64 accepted ...
Страница 80: ......